Skip to content

Exactech Equinoxe Small Reverse Shoulder System

by Exactech, Inc.

510(k) K180632

Exactech Equinoxe Small Reverse Shoulder System is a medical device (FDA Class II; FDA product code PHX, Shoulder Prosthesis, Reverse Configuration) made by Exactech, Inc., cleared under FDA 510(k) K180632 on 25 Jul 2018, active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK180632 Date25 Jul 2018ClearedStatusActive

Official records

From the FDA record

Received by the FDA
12 Mar 2018
FDA decision
25 Jul 2018 (Substantially Equivalent)
FDA review time
135 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K180632 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

3 recalls · 3 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1420-2024 FDA Class II initiated 6 Mar 2024
    • Ongoing Z-1421-2024 FDA Class II initiated 6 Mar 2024

    The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

    • Ongoing Z-0579-2024 FDA Class II initiated 21 Nov 2023 on shared 510(k) K180632

    Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

4 dated events

  1. Cleared FDA 510(k) K180632
  2. Recall initiated FDA recalls Z-0579-2024
  3. Recall initiated FDA recalls Z-1420-2024
  4. Recall initiated FDA recalls Z-1421-2024

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
PHX
Device class
Class II
Regulation
21 CFR Part 888.3660
Advisory committee
Orthopedic
FDA decision
SESE Substantially Equivalent

Devices cleared under K180632

2 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K180632
UDI-DISize and differences
Equinoxe 320-15-10
Exactech, Inc.
320-15-10 / 320-15-1010885862243997
Single use: yes
On the market
Equinoxe SPI-041516
Exactech, Inc.
SPI-041516 / SPI-04151610885862606808
Single use: no
On the market

Similar devices

Other devices filed under FDA product code PHX (Shoulder Prosthesis, Reverse Configuration).

Other devices from Exactech, Inc.

See all devices from Exactech, Inc. →