Interlink 2N3399
Interlink 2N3399 is a catheter injection port (FDA Class II; FDA product code FPA, Set, administration, intravascular) made by Baxter Healthcare Corp, cleared under FDA 510(k) K925126 on 18 Jun 1993 and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 00085412004983), active as of 23 Oct 2026. 24 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00085412004983 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK925126 | Date18 Jun 1993 | StatusActive |
Official records
- AccessGUDID record 00085412004983
- FDA 510(k) database K925126
From the FDA record
- Device description
- Injection Site
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K925126 decided 18 Jun 1993 Active
- FDA GUDID UDI-DI 00085412004983 listed 24 Sep 2016 Listed
Recalls
-
- Terminated Z-1269-2013
Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
-
- Terminated Z-0471-2012
During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.
-
- Terminated Z-2837-2011
Some of the individual extension set packages were improperly sealed, compromising the sterility of the product.
-
- Terminated Z-0629-06
- Terminated Z-0630-06
- Terminated Z-0634-06
- Terminated Z-0637-06
- Terminated Z-0638-06
- Terminated Z-0640-06
- Terminated Z-0641-06
- Terminated Z-0644-06
- Terminated Z-0648-06
- Terminated Z-0652-06
- Terminated Z-1424-05
The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
-
- Terminated Z-1099-05
- Terminated Z-1102-05
Leaks and disconnection between the bifurcated Y-site and the set tubing may occur and result in underdelivery of medication, blood loss, and a breach of the sterile fluid pathway.
-
- Terminated Z-1151-03
- Terminated Z-1154-03
- Terminated Z-1165-03
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
-
- Terminated Z-0450-03
Leakage may occur at the inlet port
-
- Terminated Z-0505-03
- Terminated Z-0507-03
- Terminated Z-0509-03
- Terminated Z-0517-03
- Terminated Z-0518-03
- Terminated Z-0522-03
The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
Timeline
- Recall initiated FDA recalls Z-0638-06
- Recall initiated FDA recalls Z-0640-06
- Recall initiated FDA recalls Z-0641-06
- Recall initiated FDA recalls Z-0644-06
- Recall initiated FDA recalls Z-0648-06
- Recall initiated FDA recalls Z-0652-06
- Recall initiated FDA recalls Z-2837-2011
- Recall initiated FDA recalls Z-0471-2012
- Recall initiated FDA recalls Z-1269-2013
- Listed FDA GUDID 00085412004983
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Interlink
- Generic name
- Catheter injection port
- Model / Ref / Catalog
- 2N3399
- Primary DI / UDI-DI
- 00085412004983
- 510(k) Number
- FDA product code
- FPA
- GMDN term
- Catheter injection port
- Device class
- Class II
- Regulation number
- 880.5440
- GUDID distribution status
- In Commercial Distribution