InterStim™ 3560030
InterStim™ 3560030 is a Class III device made by Medtronic, Inc., approved under FDA PMA P080025 on 14 Mar 2011 (1 earlier submission) and listed in FDA GUDID since 10 Aug 2020 (UDI-DI 00763000203887), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00763000203887 | Date10 Aug 2020 | StatusListed |
| United States | FDA PMAAPPR | NumberP080025 | Date14 Mar 2011 | StatusActive |
Official records
- AccessGUDID record 00763000203887
- FDA PMA database P080025
- FDA PMA database P970004
Across registries
- FDA PMA P080025 decided 14 Mar 2011 Active
- FDA GUDID UDI-DI 00763000203887 listed 10 Aug 2020 Listed
Recalls
-
- Ongoing Z-1704-2025
Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Patient may experience return of underlying disease symptoms (e.g. overactive bladder, urinary retention, or fecal incontinence) based on their response to stimulation during the evaluation period.
-
- Ongoing Z-0372-2025
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
-
- Ongoing Z-2365-2023
Lot of InterStim Self-Adhesive Ground Pad is being recalled due to the product being labeled with the incorrect expiration date.
-
- Terminated Z-0998-2022
There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.
-
- Terminated Z-1147-2015
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
-
- Terminated Z-0037-2008
A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .
-
- Terminated Z-0040-2008
Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).
Timeline
- Approved FDA PMA P970004
- Recall initiated FDA recalls Z-0037-2008
- Recall initiated FDA recalls Z-0040-2008
- Approved FDA PMA P080025
- Recall initiated FDA recalls Z-1147-2015
- Listed FDA GUDID 00763000203887
- Recall initiated FDA recalls Z-0998-2022
- Recall initiated FDA recalls Z-2365-2023
- Recall initiated FDA recalls Z-0372-2025
- Recall initiated FDA recalls Z-1704-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- InterStim™
- Generic name
- Sacral plexus percutaneous incontinence-control electrical stimulation system electrode
- Model / reference
- 3560030
- Primary DI / UDI-DI
- 00763000203887
- FDA product code
- EZW, QON
- GMDN term
- Sacral plexus percutaneous incontinence-control electrical stimulation system electrode
- Device class
- Class III
- Regulation number
- 876.5270
- GUDID distribution status
- In Commercial Distribution
- Size
- Length — 100.0 Centimeter