Interstim® 3023
Interstim® 3023 is a Class III device made by Medtronic, Inc., approved under FDA PMA P970004 on 29 Sep 1997 and listed in FDA GUDID since 23 Sep 2014 (UDI-DI 00613994287731), active as of 23 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00613994287731 | Date23 Sep 2014 | StatusListed |
| United States | FDA PMAAPPR | NumberP970004 | Date29 Sep 1997 | StatusActive |
Official records
- AccessGUDID record 00613994287731
- FDA PMA database P970004
Across registries
- FDA PMA P970004 decided 29 Sep 1997 Active
- FDA GUDID UDI-DI 00613994287731 listed 23 Sep 2014 Listed
Recalls
-
- Ongoing Z-0372-2025
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
-
- Terminated Z-0998-2022
There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.
-
- Terminated Z-1147-2015
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
-
- Terminated Z-0037-2008
A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .
-
- Terminated Z-0040-2008
Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).
Timeline
- Approved FDA PMA P970004
- Recall initiated FDA recalls Z-0037-2008
- Recall initiated FDA recalls Z-0040-2008
- Listed FDA GUDID 00613994287731
- Recall initiated FDA recalls Z-1147-2015
- Recall initiated FDA recalls Z-0998-2022
- Recall initiated FDA recalls Z-0372-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Interstim®
- Generic name
- Implantable incontinence-control electrical stimulation system pulse generator
- Model / reference
- 3023
- Primary DI / UDI-DI
- 00613994287731
- PMA number
- FDA product code
- EZW, QON
- GMDN term
- Implantable incontinence-control electrical stimulation system pulse generator
- Device class
- Class III
- Regulation number
- 876.5270
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code QON.
- Axonics Sacral Neuromodulation System by Boston Scientific
- Neuspera Sacral Neuromodulation System by NEUSPERA MEDICAL, INC.
- Interstim Therapy System for Urinary Control by MEDTRONIC NEUROMODULATION
- Virtis Sacral Neuromodulation System by Cirtec Medical Corporation
- InterStim Therapy System, Verify Evaluation System by MEDTRONIC NEUROMODULATION