Link MP Reconstruction Prosthesis
by WALDEMAR LINK GmbH & Co. KG
510(k) K142187
Link MP Reconstruction Prosthesis is a medical device (FDA Class II; FDA product code LZO) made by WALDEMAR LINK GmbH & Co. KG, cleared under FDA 510(k) K142187 on 6 Nov 2014, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK142187 | Date6 Nov 2014 | StatusActive |
Official records
- FDA 510(k) database K142187
From the FDA record
- Received by the FDA
- 8 Aug 2014
- FDA decision
- 6 Nov 2014 (Substantially Equivalent)
- FDA review time
- 90 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-0769-2025
Error on the main label and the patient label. The labels indicate a size XXL, this is a standard-size neck segment.. All further information on the labels is correct (Length, CCD and Offset).
-
- Terminated Z-0733-2019
The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
Timeline
- Cleared FDA 510(k) K142187
- Recall initiated FDA recalls Z-0733-2019
- Recall initiated FDA recalls Z-0769-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- LZO
- Device class
- Class II
- Regulation
- 21 CFR Part 888.3353
- Advisory committee
- Orthopedic
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K142187
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| MP Reconstruction Prosthesis 172-916/12 | 172-916/12 / 172-916/12 | 04026575320509 | — | On the market |
| MP Reconstruction Prosthesis 172-916/16 | 172-916/16 / 172-916/16 | 04026575320547 | — | On the market |
| MP Reconstruction Prosthesis 172-916/18 | 172-916/18 / 172-916/18 | 04026575320554 | — | On the market |
| MP Reconstruction Prosthesis 172-916/20 | 172-916/20 / 172-916/20 | 04026575320561 | — | On the market |
| MP Reconstruction Prosthesis 172-916/22 | 172-916/22 / 172-916/22 | 04026575320578 | — | On the market |
| MP Reconstruction Prosthesis 172-916/25 | 172-916/25 / 172-916/25 | 04026575320585 | — | On the market |
| MP Reconstruction Prosthesis 172-918/12 | 172-918/12 / 172-918/12 | 04026575320592 | — | On the market |
| MP Reconstruction Prosthesis 172-918/14 | 172-918/14 / 172-918/14 | 04026575320608 | — | On the market |
| MP Reconstruction Prosthesis 172-918/16 | 172-918/16 / 172-918/16 | 04026575320615 | — | On the market |
| MP Reconstruction Prosthesis 172-918/18 | 172-918/18 / 172-918/18 | 04026575320622 | — | On the market |
| MP Reconstruction Prosthesis 172-918/22 | 172-918/22 / 172-918/22 | 04026575320646 | — | On the market |
| MP Reconstruction Prosthesis 172-918/25 | 172-918/25 / 172-918/25 | 04026575320653 | — | On the market |
| MP Reconstruction Prosthesis 172-921/12 | 172-921/12 / 172-921/12 | 04026575320677 | — | On the market |
| MP Reconstruction Prosthesis 172-921/16 | 172-921/16 / 172-921/16 | 04026575210718 | — | On the market |
| MP Reconstruction Prosthesis 172-921/18 | 172-921/18 / 172-921/18 | 04026575210725 | — | On the market |
| MP Reconstruction Prosthesis 172-921/20 | 172-921/20 / 172-921/20 | 04026575210732 | — | On the market |
| MP Reconstruction Prosthesis 172-921/22 | 172-921/22 / 172-921/22 | 04026575210749 | — | On the market |
| MP Reconstruction Prosthesis 172-921/25 | 172-921/25 / 172-921/25 | 04026575210756 | — | On the market |
| MP Reconstruction Prosthesis 172-925/14 | 172-925/14 / 172-925/14 | 04026575203840 | — | On the market |
| MP Reconstruction Prosthesis 172-925/22 | 172-925/22 / 172-925/22 | 04026575203895 | — | On the market |
| MP Reconstruction Prosthesis 172-925/25 | 172-925/25 / 172-925/25 | 04026575203901 | — | On the market |
| MP Reconstruction Prosthesis 172-929/14 | 172-929/14 / 172-929/14 | 04026575202904 | — | On the market |
| MP Reconstruction Prosthesis 172-929/16 | 172-929/16 / 172-929/16 | 04026575202911 | — | On the market |
| MP Reconstruction Prosthesis 172-929/18 | 172-929/18 / 172-929/18 | 04026575202928 | — | On the market |
| MP Reconstruction Prosthesis 172-929/22 | 172-929/22 / 172-929/22 | 04026575202942 | — | On the market |
| MP Reconstruction Prosthesis 172-930/14 | 172-930/14 / 172-930/14 | 04026575202966 | — | On the market |
| MP Reconstruction Prosthesis 172-930/16 | 172-930/16 / 172-930/16 | 04026575202973 | — | On the market |
| MP Reconstruction Prosthesis 172-930/18 | 172-930/18 / 172-930/18 | 04026575202980 | — | On the market |
| MP Reconstruction Prosthesis 172-930/20 | 172-930/20 / 172-930/20 | 04026575202997 | — | On the market |
| MP Reconstruction Prosthesis 172-930/22 | 172-930/22 / 172-930/22 | 04026575203000 | — | On the market |
| MP Reconstruction Prosthesis 172-930/25 | 172-930/25 / 172-930/25 | 04026575203017 | — | On the market |
| MP Reconstruction Prosthesis 172-950/20 | 172-950/20 / 172-950/20 | 04026575203079 | — | On the market |
| MP Reconstruction Prosthesis 172-960/26 | 172-960/26 / 172-960/26 | 04026575270538 | — | On the market |
| MP Reconstruction Prosthesis 172-960/35 | 172-960/35 / 172-960/35 | 04026575270545 | — | On the market |
| MP Reconstruction Prosthesis 172-961/26 | 172-961/26 / 172-961/26 | 04026575270576 | — | On the market |
| MP Reconstruction Prosthesis 172-961/35 | 172-961/35 / 172-961/35 | 04026575270583 | — | On the market |
| MP Reconstruction Prosthesis 172-964/26 | 172-964/26 / 172-964/26 | 04026575270552 | — | On the market |
| MP Reconstruction Prosthesis 172-965/26 | 172-965/26 / 172-965/26 | 04026575270606 | — | On the market |
| MP Reconstruction Prosthesis 172-965/35 | 172-965/35 / 172-965/35 | 04026575270613 | — | On the market |
| MP Reconstruction Prosthesis 172-980/35 | 172-980/35 / 172-980/35 | 04026575297283 | — | On the market |
| MP Reconstruction Prosthesis 172-981/26 | 172-981/26 / 172-981/26 | 04026575297313 | — | On the market |
| MP Reconstruction Prosthesis 172-981/35 | 172-981/35 / 172-981/35 | 04026575297320 | — | On the market |
| MP Reconstruction Prosthesis 172-984/26 | 172-984/26 / 172-984/26 | 04026575297290 | — | On the market |
| MP Reconstruction Prosthesis 172-984/35 | 172-984/35 / 172-984/35 | 04026575297306 | — | On the market |
| MP Reconstruction Prosthesis 172-985/26 | 172-985/26 / 172-985/26 | 04026575297337 | — | On the market |
| MP Reconstruction Prosthesis 172-985/35 | 172-985/35 / 172-985/35 | 04026575297344 | — | On the market |
| MP Reconstruction Prosthesis 99-0984/28 | 99-0984/28 / 99-0984/28 | 04026575409747 | — | On the market |
| MP Reconstruction Prosthesis 99-0984/30 | 99-0984/30 / 99-0984/30 | 04026575409761 | — | On the market |
| MP Reconstruction Prosthesis 99-0984/32 | 99-0984/32 / 99-0984/32 | 04026575409785 | — | On the market |
| MP Reconstruction System 172-916/14 | 172-916/14 | 04026575320530 | — | On the market |
Similar devices
Other devices filed under FDA product code LZO (Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented).
- MiniHip Stem 580.90239999999994 by CORIN LTD
- 22.2/+4MM CO CR & ZIRCONIA FEMORAL HEAD by Landos, Inc.
- 65-7711-015-00 by Zimmer, Inc.
- Trabecular Metal® 6202-68-22 00-6202-068-22 by Zimmer, Inc.
- MiniHip Stem 580.9604 by CORIN LTD
- G7® 010002722 by Biomet Orthopedics, Inc.
- MASTERSL Uncemented Stem 3516.21.330 by Limacorporate
- CORIN MINIHIP FEMORAL STEM by Corin USA
- G7® 010002457 by Biomet Orthopedics, Inc.
- APR OVERSIZED HIP STEM by Intermedics Orthopedics