Skip to content

Luminos Agile

by Siemens Medical Solutions USA, Inc.

510(k) K111292

Luminos Agile is a medical device (FDA Class II; FDA product code JAA, System, X-Ray, Fluoroscopic, Image-Intensified) made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K111292 on 3 Jun 2011, active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK111292 Date3 Jun 2011ClearedStatusActive

Official records

From the FDA record

Received by the FDA
6 May 2011
FDA decision
3 Jun 2011 (Substantially Equivalent)
FDA review time
28 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K111292 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

6 recalls · 1 ongoing · FDA

    • Ongoing Z-1001-2023 FDA Class II initiated 9 Dec 2022

    There is potential for the footrest to detach from the patient table during use.

    • Terminated Z-1669-2015 FDA Class II initiated 13 Apr 2015

    Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

    • Terminated Z-0994-2015 FDA Class II initiated 15 Dec 2014 on shared 510(k) K111292

    Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.

    • Terminated Z-0005-2015 FDA Class II initiated 2 Sep 2014 on shared 510(k) K111292

    Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

    • Terminated Z-0374-2014 FDA Class II initiated 3 Sep 2013 on shared 510(k) K111292

    A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.

    • Terminated Z-1159-2012 FDA Not yet classified initiated 15 Feb 2012 on shared 510(k) K111292

    Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

7 dated events

  1. Cleared FDA 510(k) K111292
  2. Recall initiated FDA recalls Z-1159-2012
  3. Recall initiated FDA recalls Z-0374-2014
  4. Recall initiated FDA recalls Z-0005-2015
  5. Recall initiated FDA recalls Z-0994-2015
  6. Recall initiated FDA recalls Z-1669-2015
  7. Recall initiated FDA recalls Z-1001-2023

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
JAA
Device class
Class II
Regulation
21 CFR Part 892.1650
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code JAA (System, X-Ray, Fluoroscopic, Image-Intensified).

Other devices from Siemens Medical Solutions USA, Inc.

See all devices from Siemens Medical Solutions USA, Inc. →