Luminos Agile
by Siemens Medical Solutions USA, Inc.
510(k) K111292
Luminos Agile is a medical device (FDA Class II; FDA product code JAA, System, X-Ray, Fluoroscopic, Image-Intensified) made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K111292 on 3 Jun 2011, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK111292 | Date3 Jun 2011 | StatusActive |
Official records
- FDA 510(k) database K111292
From the FDA record
- Received by the FDA
- 6 May 2011
- FDA decision
- 3 Jun 2011 (Substantially Equivalent)
- FDA review time
- 28 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1001-2023
There is potential for the footrest to detach from the patient table during use.
-
- Terminated Z-1669-2015
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
-
- Terminated Z-0994-2015
Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.
-
- Terminated Z-0005-2015
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
-
- Terminated Z-0374-2014
A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.
-
- Terminated Z-1159-2012
Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.
Timeline
- Cleared FDA 510(k) K111292
- Recall initiated FDA recalls Z-1159-2012
- Recall initiated FDA recalls Z-0374-2014
- Recall initiated FDA recalls Z-0005-2015
- Recall initiated FDA recalls Z-0994-2015
- Recall initiated FDA recalls Z-1669-2015
- Recall initiated FDA recalls Z-1001-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAA
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JAA (System, X-Ray, Fluoroscopic, Image-Intensified).
- Uroskop Omnia Max 10762473 by Siemens Healthcare GmbH
- LIXI IMAGING SCOPE. MODIFICATION by Lixi, Inc.
- FLUOROSCAN/TM BY HEALTHMATE by Healthmate, Inc.
- HANEL MAVIG MONITOR SUSPENSION SYSTEM by F. Walter Hanel GmbH
- TELSTAR BIPLANE DIGITAL IMAGING SYSTEM [TIS] MODEL 030-000915 by Stereotaxis, Inc.
- UNBRANDED 07-CA201 by Advance Medical Designs, Inc.
- CIOS SELECT 11515088 by Siemens Shanghai Medical Equipment Ltd.
- JOINT MOTON IMAGER(TM) by Concord Imaging, Ltd.
- CS-series-FP Radiographic / Fluoroscopic Systems with Optional CA-100S / FluoroShield™ (ROI Assembly) by Omega Medical Imaging, LLC
- UNBRANDED 03-KP30 by Advance Medical Designs, Inc.