by Siemens Medical Solutions USA, Inc.

510(k) K062623

Axiom Luminos DRF is a medical device (FDA Class II; FDA product code OWB, Interventional Fluoroscopic X-Ray System) made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K062623 on 22 Aug 2007, active as of 22 Oct 2026. 7 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK062623 Date22 Aug 2007ClearedStatusActive

Official records

From the FDA record

Received by the FDA
5 Sep 2006
FDA decision
22 Aug 2007 (Substantially Equivalent)
FDA review time
351 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K062623 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

20 recalls · 2 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1164-2023 FDA Class II initiated 21 Dec 2022
    • Ongoing Z-1166-2023 FDA Class II initiated 21 Dec 2022

    Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values

    • Terminated Z-1668-2015 FDA Class II initiated 13 Apr 2015

    Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

    • Terminated Z-0994-2015 FDA Class II initiated 15 Dec 2014 on shared 510(k) K062623

    Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.

    • Terminated Z-0003-2015 FDA Class II initiated 2 Sep 2014 on shared 510(k) K062623

    Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

    • Terminated Z-2649-2014 FDA Class II initiated 12 Aug 2014

    It was determined that a potential malfunction may occur on the Luminos dRF system with SW VD10 or on Luminos dRF Max when the table side control is exposed to fluids. Fluid may infiltrate the tableside control panel and result in malfunction. In the very unlikely case of fluid infiltration a permanent short circuit within the control panel of the Luminos dRF / Luminos dRF Max will automatically d

    • Terminated Z-2292-2014 FDA Class II initiated 7 Jul 2014 on shared 510(k) K062623

    It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. Thi

    • Terminated Z-1711-2014 FDA Class II initiated 11 Apr 2014

    It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit.

    • Terminated Z-0140-2014 FDA Class II initiated 3 Sep 2013 on shared 510(k) K062623

    When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.

    • Terminated Z-0374-2014 FDA Class II initiated 3 Sep 2013 on shared 510(k) K062623

    A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.

    • Terminated Z-1845-2014 FDA Class II initiated 28 Aug 2013 on shared 510(k) K062623

    When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.

    • Terminated Z-1848-2013 FDA Class II initiated 7 May 2013

    Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.

    • Terminated Z-0016-2014 FDA Class II initiated 6 May 2013

    Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition. Additionally, it is not possible for the user to stop the radiation manually; ho

    • Terminated Z-1383-2012 FDA Not yet classified initiated 16 Feb 2012 on shared 510(k) K062623

    During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250

    • Terminated Z-1159-2012 FDA Not yet classified initiated 15 Feb 2012 on shared 510(k) K062623

    Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.

    • Terminated Z-1421-2011 FDA Unclassified (pre-2012) initiated 14 Jan 2011 on shared 510(k) K062623

    Firm became aware of a potential malfunction and possible hazard to patients in the event the tilt buttons on the OPTI Grip become stuck under the housing

    • Terminated Z-0717-2011 FDA Unclassified (pre-2012) initiated 25 Oct 2010 on shared 510(k) K062623

    patient leg can be pinched

    • Terminated Z-2222-2010 FDA Unclassified (pre-2012) initiated 13 Jul 2010 on shared 510(k) K062623

    Fluid may enter the multi-functional handle (Opti Grip) when using the Axiom Luminos TF or Sireskop SD system, causing unintended reaction or movement, increased radiation or useless images.

  • 2 recalls, same reason

    • Terminated Z-0327-2009 FDA Unclassified (pre-2012) initiated 19 Sep 2008 on shared 510(k) K062623
    • Terminated Z-0326-2009 FDA Unclassified (pre-2012) initiated 19 Sep 2008 on shared 510(k) K062623

    Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

21 dated events

  1. Recall initiated FDA recalls Z-0140-2014
  2. Recall initiated FDA recalls Z-0374-2014
  3. Recall initiated FDA recalls Z-1711-2014
  4. Recall initiated FDA recalls Z-2292-2014
  5. Recall initiated FDA recalls Z-2649-2014
  6. Recall initiated FDA recalls Z-0003-2015
  7. Recall initiated FDA recalls Z-0994-2015
  8. Recall initiated FDA recalls Z-1668-2015
  9. Recall initiated FDA recalls Z-1164-2023
  10. Recall initiated FDA recalls Z-1166-2023

11 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
OWB
Device class
Class II
Regulation
21 CFR Part 892.1650
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code OWB (Interventional Fluoroscopic X-Ray System).