Magnetom Trio System
by Siemens Medical Systems, Inc.
510(k) K013586
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK013586 | Date28 Dec 2001 | StatusActive |
Recalls
-
- Terminated Z-2021-2011
Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
-
- Terminated Z-1404-04
- Terminated Z-1403-04
An input error can be made with Magnetom MRI software. Software allows users to manually change Transmit Reference Voltage in certain windows
-
- Terminated Z-0521-04
tissue roll can become magnetic
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent