FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK030531 Date17 Jun 2003ClearedStatusActive

Recalls

3 recalls · FDA

    • Terminated Z-1007-2007 FDA Unclassified (pre-2012) initiated 27 Apr 2007

    Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi

    • Terminated Z-0010-05 FDA Unclassified (pre-2012) initiated 14 Sep 2004

    Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The investigation revealed this only occurs when the pump is programmed to display information in Spanish.

    • Terminated Z-0763-04 FDA Unclassified (pre-2012) initiated 29 Jan 2004 on shared 510(k) K030531

    39 lots of BD Test strip may produce an increased frequency of E-3 messages when used during blood glucose testing.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LZG
Device Class
Class II
Regulation
21 CFR Part 880.5725
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent