Minimed Paradigm Insulin Pump, Model 511
510(k) K001829
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK001829 | Date18 Aug 2000 | StatusActive |
Recalls
-
- Terminated Z-0178-2020
There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps.
-
- Terminated Z-1007-2007
Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi
-
- Terminated Z-0168-04
Water ingresses into the device when in prolonged contact with water causing device to malfunction.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent