Skip to content

Mevatron 67

by Siemens Medical Laboratories, Inc.

510(k) K790688

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK790688 Date23 Apr 1979ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-1433-2017 FDA Class II initiated 25 Jan 2017

    Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.

    • Terminated Z-2174-2014 FDA Class II initiated 28 May 2014

    A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYE
Device Class
Class II
Regulation
21 CFR Part 892.5050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Siemens Medical Laboratories, Inc.