Mevatron Kd-2

by Siemens Medical Laboratories, Inc.

510(k) K862339

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

16 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK862339 Date29 Oct 1986ClearedStatusActive

Recalls

16 recalls · FDA

    • Terminated Z-1437-2017 FDA Class II initiated 25 Jan 2017

    Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.

  • 3 recalls, same reason

    • Terminated Z-2636-2011 FDA Unclassified (pre-2012) initiated 4 Nov 2010
    • Terminated Z-2638-2011 FDA Unclassified (pre-2012) initiated 4 Nov 2010
    • Terminated Z-2637-2011 FDA Unclassified (pre-2012) initiated 4 Nov 2010

    The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.

    • Terminated Z-1395-2008 FDA Unclassified (pre-2012) initiated 11 Jan 2008

    Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.

  • 5 recalls, same reason

    • Terminated Z-1756-2008 FDA Unclassified (pre-2012) initiated 10 Dec 2007
    • Terminated Z-1753-2008 FDA Unclassified (pre-2012) initiated 10 Dec 2007
    • Terminated Z-1757-2008 FDA Unclassified (pre-2012) initiated 10 Dec 2007
    • Terminated Z-1755-2008 FDA Unclassified (pre-2012) initiated 10 Dec 2007
    • Terminated Z-1758-2008 FDA Unclassified (pre-2012) initiated 10 Dec 2007

    Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.

    • Terminated Z-1754-2008 FDA Unclassified (pre-2012) initiated 10 Dec 2007

    Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.

  • 4 recalls, same reason

    • Terminated Z-0685-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0680-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0677-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0681-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007

    Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.

    • Terminated Z-0851-2008 FDA Unclassified (pre-2012) initiated 27 Jul 2007

    Door Hinge failures; loose or falling off (Stationary Structure Doors)

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYE
Device Class
Class II
Regulation
21 CFR Part 892.5050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent