Mevatron Kd-2
by Siemens Medical Laboratories, Inc.
510(k) K862339
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK862339 | Date29 Oct 1986 | StatusActive |
Recalls
-
- Terminated Z-1437-2017
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
-
- Terminated Z-2636-2011
- Terminated Z-2638-2011
- Terminated Z-2637-2011
The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
-
- Terminated Z-1395-2008
Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
-
- Terminated Z-1756-2008
- Terminated Z-1753-2008
- Terminated Z-1757-2008
- Terminated Z-1755-2008
- Terminated Z-1758-2008
Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
-
- Terminated Z-1754-2008
Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
-
- Terminated Z-0685-2008
- Terminated Z-0680-2008
- Terminated Z-0677-2008
- Terminated Z-0681-2008
Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
-
- Terminated Z-0851-2008
Door Hinge failures; loose or falling off (Stationary Structure Doors)
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYE
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent