Mevatron, M 2 Class
by Siemens Medical Laboratories, Inc.
510(k) K882729
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK882729 | Date19 Sep 1988 | StatusActive |
Recalls
-
- Terminated Z-1679-2020
- Terminated Z-1678-2020
Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
-
- Terminated Z-1436-2017
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
-
- Terminated Z-0790-2012
- Terminated Z-0794-2012
- Terminated Z-0797-2012
- Terminated Z-0792-2012
- Terminated Z-0781-2012
- Terminated Z-0788-2012
- Terminated Z-0793-2012
- Terminated Z-0798-2012
- Terminated Z-0795-2012
- Terminated Z-0787-2012
- Terminated Z-0786-2012
- Terminated Z-0780-2012
- Terminated Z-0785-2012
- Terminated Z-0789-2012
- Terminated Z-0791-2012
- Terminated Z-0783-2012
- Terminated Z-0784-2012
- Terminated Z-0796-2012
Bolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover.
-
- Terminated Z-1402-2008
- Terminated Z-1401-2008
Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
-
- Terminated Z-0679-2008
- Terminated Z-0682-2008
- Terminated Z-0684-2008
- Terminated Z-0678-2008
- Terminated Z-0683-2008
Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
-
- Terminated Z-0852-2008
Door Hinge failures; loose or falling off (Stationary Structure Doors)
-
- Terminated Z-0156-2008
- Terminated Z-0157-2008
- Terminated Z-0159-2008
- Terminated Z-0158-2008
- Terminated Z-0163-2008
- Terminated Z-0160-2008
- Terminated Z-0155-2008
- Terminated Z-0162-2008
- Terminated Z-0161-2008
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment plans. A second reason is an unplanned Dosimetry calibration data set can be erroneously copied during "Service Mode."
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYE
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent