Skip to content

Mevatron, M 2 Class

by Siemens Medical Laboratories, Inc.

510(k) K882729

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

38 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK882729 Date19 Sep 1988ClearedStatusActive

Recalls

38 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-1679-2020 FDA Class II initiated 18 Feb 2020
    • Terminated Z-1678-2020 FDA Class II initiated 18 Feb 2020

    Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.

    • Terminated Z-1436-2017 FDA Class II initiated 25 Jan 2017

    Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.

  • 18 recalls, same reason

    • Terminated Z-0790-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0794-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0797-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0792-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0781-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0788-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0793-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0798-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0795-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0787-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0786-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0780-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0785-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0789-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0791-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0783-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0784-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011
    • Terminated Z-0796-2012 FDA Unclassified (pre-2012) initiated 1 Dec 2011

    Bolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover.

  • 2 recalls, same reason

    • Terminated Z-1402-2008 FDA Unclassified (pre-2012) initiated 11 Jan 2008
    • Terminated Z-1401-2008 FDA Unclassified (pre-2012) initiated 11 Jan 2008

    Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.

  • 5 recalls, same reason

    • Terminated Z-0679-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0682-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0684-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0678-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007
    • Terminated Z-0683-2008 FDA Unclassified (pre-2012) initiated 7 Aug 2007

    Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.

    • Terminated Z-0852-2008 FDA Unclassified (pre-2012) initiated 27 Jul 2007

    Door Hinge failures; loose or falling off (Stationary Structure Doors)

  • 9 recalls, same reason

    • Terminated Z-0156-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0157-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0159-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0158-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0163-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0160-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0155-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0162-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007
    • Terminated Z-0161-2008 FDA Unclassified (pre-2012) initiated 22 Mar 2007

    Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment plans. A second reason is an unplanned Dosimetry calibration data set can be erroneously copied during "Service Mode."

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYE
Device Class
Class II
Regulation
21 CFR Part 892.5050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Siemens Medical Laboratories, Inc.

See all devices from Siemens Medical Laboratories, Inc. →