microTargeting™ Frame Adapter 66-FA-LX-00391
by Fhc, Inc.
microTargeting Frame Adapter 66-FA-LX-00391, a neurosurgical micro targeting drive (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Fhc, Inc., was cleared under FDA 510(k) K011992 on 14 Aug 2001 and listed in FDA GUDID on 17 Sep 2016 (UDI-DI 00873263003916); the registry lists it as not in commercial distribution (GUDID) as of 22 Oct 2026. FDA recalls on record: 6.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00873263003916 | Date17 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK011992 | Date14 Aug 2001 | StatusActive |
Official records
- AccessGUDID record 00873263003916
- FDA 510(k) database K011992
From the FDA record
- Device description
- Leksell Frame Adapter for use with microTargeting Drive
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use
Across registries
- FDA 510(k) K011992 decided 14 Aug 2001 Active
- FDA GUDID UDI-DI 00873263003916 listed 17 Sep 2016 Listed
Recalls
-
- Terminated Z-2353-2018
- Terminated Z-2359-2018
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
-
- Terminated Z-0486-2009
Measuring fixture is incorrectly graduated.
-
- Terminated Z-0179-06
- Terminated Z-0180-06
Power Assist microTargeting System may advance beyond intended target
-
- Terminated Z-0895-05
Potential for non-sterile pin to contaminate sterile field
Timeline
- Cleared FDA 510(k) K011992
- Recall initiated FDA recalls Z-0895-05
- Recall initiated FDA recalls Z-0179-06
- Recall initiated FDA recalls Z-0180-06
- Recall initiated FDA recalls Z-0486-2009
- Listed FDA GUDID 00873263003916
- Recall initiated FDA recalls Z-2353-2018
- Recall initiated FDA recalls Z-2359-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- microTargeting™ Frame Adapter
- Generic name
- Neurosurgical micro targeting drive
- Model / Ref / Catalog
- 66-FA-LX-00391, 66-FA-LX
- Primary DI / UDI-DI
- 00873263003916
- 510(k) Number
- FDA product code
- HAW
- GMDN term
- Neurosurgical micro targeting drive
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- Not in Commercial Distribution
Similar devices
Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).
- STEALTHSTATION SYSTEM by Surgical Navigation Technologies, Inc.
- KOLIBRI IGS SYSTEMS by Brainlab AG
- RENAISSANCE SYSTEM by MAZOR ROBOTICS LTD
- VECTORVISION HIP by Brainlab AG
- STYLET, 23CM, SKULL-MOUNT PATIENT TRACKER, NON-INVASIVE PATIENT TRACKER, TRACER POINTER, TOUCH-N-GO POINTER, NAVIGATION by Medtronic Navigation Inc.
- StealthStation S8 9735667 by Medtronic Navigation Inc.
- PIGALILEO CAS, VERSION 4.0 AND TKR BASE, VERSION 2.1 by Smith & Nephew, Inc.
- Zeta Cranial Navigation System (ZNS131-US) by ZETA SURGICAL INC.
- PHILIPS EASYGUIDE by Philips Medical Systems
- 3D LINE STEROTACTIC HARDWARE ACCESSORIES by 3D Line USA, Inc.