microTargeting™ Insertion Tube 66-IT(FP3)-00664
by Fhc, Inc.
microTargeting Insertion Tube 66-IT(FP3)-00664 is an intracerebral cannula (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Fhc, Inc., cleared under FDA 510(k) K011775 on 30 Aug 2001 and listed in FDA GUDID since 18 May 2018 (UDI-DI 00873263006641), active as of 23 Oct 2026. 8 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00873263006641 | Date18 May 2018 | StatusListed |
| United States | FDA 510(k)SESE | NumberK011775 | Date30 Aug 2001 | StatusActive |
Official records
- AccessGUDID record 00873263006641
- FDA 510(k) database K011775
From the FDA record
- Device description
- microTargeting™ Tandem Insertion Tube, nonsterile
- MRI safety
- MR Unsafe
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use
Across registries
- FDA 510(k) K011775 decided 30 Aug 2001 Active
- FDA GUDID UDI-DI 00873263006641 listed 18 May 2018 Listed
Recalls
-
- Ongoing Z-0312-2025
A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
-
- Terminated Z-2345-2018
- Terminated Z-2346-2018
- Terminated Z-2347-2018
- Terminated Z-2348-2018
- Terminated Z-2355-2018
- Terminated Z-2356-2018
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
-
- Terminated Z-0895-05
Potential for non-sterile pin to contaminate sterile field
Timeline
- Cleared FDA 510(k) K011775
- Recall initiated FDA recalls Z-0895-05
- Recall initiated FDA recalls Z-2345-2018
- Recall initiated FDA recalls Z-2346-2018
- Recall initiated FDA recalls Z-2347-2018
- Recall initiated FDA recalls Z-2348-2018
- Recall initiated FDA recalls Z-2355-2018
- Recall initiated FDA recalls Z-2356-2018
- Listed FDA GUDID 00873263006641
- Recall initiated FDA recalls Z-0312-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- microTargeting™ Insertion Tube
- Generic name
- Intracerebral cannula, nonimplantable, reusable
- Model / Ref / Catalog
- 66-IT(FP3)-00664, 66-IT(FP3)
- Primary DI / UDI-DI
- 00873263006641
- 510(k) Number
- FDA product code
- HAW
- GMDN term
- Intracerebral cannula, nonimplantable, reusable
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Unsafe
- Size
- Lumen/Inner Diameter — 1.38 MillimeterOuter Diameter — 1.83 MillimeterLength — 211 Millimeter
Similar devices
Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).
- Stealth Autoguide™ System 28322 by Medtronic Navigation Inc.
- StealthStation System by Surgical Navigation Technologies, Inc.
- Compass CT Stereotactic Adaptation System by Stereotactic Medical Systems, Inc.
- 6000-008-000 by Stryker Leibinger
- neuro inspire by Rensihaw Plc
- 6003-250-000 by Stryker Leibinger
- Ball Joint Guide Array (66295) by Bayer Medical Care Inc.
- NavSuite3 Kit by Stryker Leibinger GmbH & Co. KG - Navigation
- NAVITRACK SYSTEM - S&N IMAGE FREE KNEE by Orthosoft Inc
- ROSA BRAIN (v3.0.0.5) by Medtech S.A