Skip to content

microTargeting™ Stylet 66-ST-14P-00237

by Fhc, Inc.

microTargeting Stylet 66-ST-14P-00237, an intracerebral cannula (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Fhc, Inc., was cleared under FDA 510(k) K011775 on 30 Aug 2001 and listed in FDA GUDID on 17 Sep 2016 (UDI-DI 00873263002377); the registry lists it as not in commercial distribution (GUDID) as of 22 Oct 2026. FDA recalls on record: 8.

Active United States. Also listed in GUDID (US), not in commercial distribution.

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

510(k) K011775

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00873263002377 UDI-DI Date17 Sep 2016ListedStatusListed GUDID: not in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK011775 Date30 Aug 2001DecidedStatusActive

Official records

From the FDA record

Device description
microTargeting DP Stylet, 1x, nonsterile, for use with mT Drive and STar Drive
MRI safety
MR Unsafe
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile; sterilise before use

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K011775 decided 30 Aug 2001 Active
  • FDA GUDID UDI-DI 00873263002377 listed 17 Sep 2016 Listed

Recalls

8 recalls · 1 ongoing · FDA

    • Ongoing Z-0312-2025 FDA Class I initiated 4 Oct 2024

    A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

  • 6 recalls, same reason

    • Terminated Z-2345-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2346-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2347-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2348-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2355-2018 FDA Class II initiated 10 May 2018
    • Terminated Z-2356-2018 FDA Class II initiated 10 May 2018

    The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

    • Terminated Z-0895-05 FDA Unclassified (pre-2012) initiated 14 Mar 2005

    Potential for non-sterile pin to contaminate sterile field

Timeline

10 dated events

  1. Cleared FDA 510(k) K011775
  2. Recall initiated FDA recalls Z-0895-05
  3. Listed FDA GUDID 00873263002377
  4. Recall initiated FDA recalls Z-2345-2018
  5. Recall initiated FDA recalls Z-2346-2018
  6. Recall initiated FDA recalls Z-2347-2018
  7. Recall initiated FDA recalls Z-2348-2018
  8. Recall initiated FDA recalls Z-2355-2018
  9. Recall initiated FDA recalls Z-2356-2018
  10. Recall initiated FDA recalls Z-0312-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
microTargeting™ Stylet
Generic name
Intracerebral cannula, implantable
Model / Ref / Catalog
66-ST-14P-00237, 66-ST-14P
Primary DI / UDI-DI
00873263002377
510(k) Number
FDA product code
HAW
GMDN term
Intracerebral cannula, implantable
Device class
Class II
Regulation number
882.4560
GUDID distribution status
Not in Commercial Distribution
MRI safety
MR Unsafe
Size
Outer Diameter — 1.83 Millimeter
Length — 204 Millimeter

Similar devices

Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).

Other devices from Fhc, Inc.

See all devices from Fhc, Inc. →

Properties

  • Supplied sterile
  • Sterilise before use
  • Single use
  • Implantable
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (8, all answered No)