Midas Rex Legend 15MH22L
by Medtronic Powered Surgical Solutions
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00763000453565 | Date12 Mar 2022 | StatusListed |
| United States | FDA 510(k)SESE | NumberK170312 | Date15 May 2017 | StatusActive |
Recalls
-
- Terminated Z-1602-2016
Incorrect color coding and attachment reference were included on the labeling of 10BA40D dissecting tools with lot number H5237248.
-
- Terminated Z-0321-2016
The wire pass hole is not present on the tools.
-
- Terminated Z-0546-2010
Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use.
-
- Terminated Z-2156-2008
Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.
-
- Terminated Z-1574-05
Potential compromise of sterile packaging for disposable dissecting tools.
-
- Terminated Z-1067-05
Complaints of finger control component breaking off and falling into open surgical wound.
-
- Terminated Z-0278-05
Softer hose increases the propensity for outer exhaust hose to kink or become occluded as a result of clamping. Complaints of hose rupture/burst have been reported.
-
- Terminated Z-0213-05
Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke).
-
- Terminated Z-0419-03
Lubricant leaking from motor after sterilization causing bleed-through sterilization wrap and/or excess lubricant in instrument case.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Midas Rex Legend
- Generic name
- Orthopaedic bur, single-use
- Model / reference
- 15MH22L
- Primary DI / UDI-DI
- 00763000453565
- FDA product code
- HBB
- GMDN term
- Orthopaedic bur, single-use
- Device class
- Class II
- Regulation number
- 882.4370
- GUDID distribution status
- In Commercial Distribution
Other devices from Medtronic Powered Surgical Solutions
Properties
Other properties (11)
- Sterilise before use: No
- Implantable: No
- Contains latex: No
- Over-the-counter: No
- Convenience kit: No
- Combination product: No
- HCT/P: No
- Premarket exempt: No
- Serial number: No
- Expiration date: No