MiniMed Flex pump
510(k) K253743
MiniMed Flex pump is a medical device (FDA Class II; FDA product code QFG, Alternate Controller Enabled Insulin Infusion Pump) made by Medtronic Minimed, Inc., cleared under FDA 510(k) K253743 on 12 Mar 2026, active as of 22 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK253743 | Date12 Mar 2026 | StatusActive |
Official records
- FDA 510(k) database K253743
From the FDA record
- Received by the FDA
- 24 Nov 2025
- FDA decision
- 12 Mar 2026 (Substantially Equivalent)
- FDA review time
- 108 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-0088-2027
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
Timeline
- Cleared FDA 510(k) K253743
- Recall initiated FDA recalls Z-0088-2027
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- QFG
- Device class
- Class II
- Regulation
- 21 CFR Part 880.5730
- Advisory committee
- Clinical Chemistry
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K253743
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| MiniMed Extended Supply Kit MMT-445B | MMT-445B | 00199150084050 | On the market | |
| MiniMed Extended Supply Kit MMT-445C | MMT-445C | 00199150084067 | On the market | |
| MiniMed Flex MMT-8062 | MMT-8062 | 00199150077380 | On the market | |
| MiniMed Flex MMT-8062 | MMT-8062 | 00763000848149 | On the market | |
| MiniMed Flex MMT-8070 | MMT-8070 | 00199150021031 | On the market | |
| MiniMed Flex MMT-8072 | MMT-8072 | 00763000848156 | On the market | |
| MiniMed Flex Starter Kit MMT-8076B | MMT-8076B | 00199150084036 | On the market | |
| MiniMed Flex Starter Kit MMT-8076C | MMT-8076C | 00199150084043 | On the market |
Similar devices
Other devices filed under FDA product code QFG (Alternate Controller Enabled Insulin Infusion Pump).
- DEKA ACE Pump System by Deka Research and Development
- t:slim Mobile Application, iOS 1014707 by Tandem Diabetes Care, Inc.
- Tandem Mobi insulin pump with interoperable technology by Tandem Diabetes Care, Inc.
- Accu-Chek Solo Micropump System with Interoperable Technology by Roche Diabetes Care GmbH
- Omnipod 5 ACE Pump (Pod) by Insulet Corporation
- iLet ACE Pump by Beta Bionics, Inc.
- twiist system by DEKA RESEARCH & DEVELOPMENT CORP.
- iLet® ACE Pump by Beta Bionics, Inc.
- Omnipod 5 ACE Pump by Insulet Corporation
- Embecta Insulin Delivery System by embecta medical I LLC