Skip to content

Modification to ABL800 Flex

by Radiometer Medical ApS

510(k) K050869

Modification to ABL800 Flex is a medical device (FDA Class II; FDA product code CIG, Diazo Colorimetry, Bilirubin) made by Radiometer Medical ApS, cleared under FDA 510(k) K050869 on 11 Aug 2005, active as of 22 Oct 2026.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK050869 Date11 Aug 2005ClearedStatusActive

Official records

From the FDA record

Received by the FDA
6 Apr 2005
FDA decision
11 Aug 2005 (Substantially Equivalent)
FDA review time
127 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K050869 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

1 recall · FDA

    • Terminated Z-0115-2007 FDA Unclassified (pre-2012) initiated 6 Jul 2006 on shared 510(k) K050869

    Incorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

2 dated events

  1. Cleared FDA 510(k) K050869
  2. Recall initiated FDA recalls Z-0115-2007

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
CIG
Device class
Class II
Regulation
21 CFR Part 862.1110
Advisory committee
Clinical Chemistry
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code CIG (Diazo Colorimetry, Bilirubin).

Other devices from Radiometer Medical ApS

See all devices from Radiometer Medical ApS →