Skip to content

Modified Distal Radius Plating System

by Biomet

510(k) K133939

Modified Distal Radius Plating System is a medical device (FDA Class II; FDA product code HRS, Plate, Fixation, Bone) made by Biomet, cleared under FDA 510(k) K133939 on 1 Apr 2014, active as of 22 Oct 2026.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK133939 Date1 Apr 2014ClearedStatusActive

Official records

From the FDA record

Received by the FDA
23 Dec 2013
FDA decision
1 Apr 2014 (Substantially Equivalent)
FDA review time
99 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K133939 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

2 recalls · FDA

    • Terminated Z-0810-2022 FDA Class II initiated 14 Feb 2022 on shared 510(k) K133939

    Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

    • Terminated Z-1082-2016 FDA Class II initiated 29 Jan 2016 on shared 510(k) K133939

    Components of the DVR ePAKs can oxidize during shipment and storage prior to use. A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. If oxidative material would flake off a component contained within the ePAK and enter the surgical field/wound an inflammatory reaction may occur. The inflammatory reaction may result in a foreign body reaction and/or metal hypersensitivity. Oxidized components have failed cytotoxicity testing.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

3 dated events

  1. Cleared FDA 510(k) K133939
  2. Recall initiated FDA recalls Z-1082-2016
  3. Recall initiated FDA recalls Z-0810-2022

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
HRS
Device class
Class II
Regulation
21 CFR Part 888.3030
Advisory committee
Orthopedic
FDA decision
SESE Substantially Equivalent

Devices cleared under K133939

14 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K133939
UDI-DISize and differences
DVR® CROSSLOCK 1318-10-051
Biomet Orthopedics, Inc.
131810051 / 1318-10-05100887868446115 —On the market
DVR® CROSSLOCK 1318-12-126
Biomet Orthopedics, Inc.
131812126 / 1318-12-12600887868446191 —On the market
DVR® CROSSLOCK 1318-12-176
Biomet Orthopedics, Inc.
131812176 / 1318-12-17600887868446207 —On the market
DVR® CROSSLOCK 1318-20-051
Biomet Orthopedics, Inc.
131820051 / 1318-20-05100887868446245 —On the market
DVR® CROSSLOCK 1318-22-126
Biomet Orthopedics, Inc.
131822126 / 1318-22-12600887868446313 —On the market
DVR® CROSSLOCK 1318-22-176
Biomet Orthopedics, Inc.
131822176 / 1318-22-17600887868446320 —On the market
DVR® CROSSLOCKEXTRA EXTRA LONG PLATE LEFT (HEAD WIDTH X OVERALL LENGTH) 1318-22-176
Biomet Orthopedics, Inc.
131822176 / 1318-22-17600880304811898 —On the market
DVR® CROSSLOCKEXTRA EXTRA LONG PLATE RIGHT (HEAD WIDTH X OVERALL LENGTH) 1318-12-176
Biomet Orthopedics, Inc.
131812176 / 1318-12-17600880304811881 —On the market
DVR® CROSSLOCKEXTRA LONG PLATE LEFT (HEAD WIDTH X OVERALL LENGTH) 1318-22-126
Biomet Orthopedics, Inc.
131822126 / 1318-22-12600880304811874 —On the market
DVR® CROSSLOCKEXTRA LONG PLATE RIGHT (HEAD WIDTH X OVERALL LENGTH) 1318-12-126
Biomet Orthopedics, Inc.
131812126 / 1318-12-12600880304811867 —On the market
DVR® CROSSLOCKEXTRA NARROW PLATE LEFT (HEAD WIDTH X OVERALL LENGTH) 1318-20-051
Biomet Orthopedics, Inc.
131820051 / 1318-20-05100880304811850 —On the market
DVR® CROSSLOCKEXTRA NARROW PLATE RIGHT (HEAD WIDTH X OVERALL LENGTH) 1318-10-051
Biomet Orthopedics, Inc.
131810051 / 1318-10-05100880304811843 —On the market
DVR® CROSSLOCKPLATE, WIDE LEFT WITH SMOOTH PEGS 811923050
Biomet Orthopedics, Inc.
811923050 / 81192305000887868129544 —On the market
DVR® CROSSLOCKPLATE, WIDE RIGHT WITH SMOOTH PEGS 811913050
Biomet Orthopedics, Inc.
811913050 / 81191305000887868129247 —On the market

Similar devices

Other devices filed under FDA product code HRS (Plate, Fixation, Bone).

Other devices from Biomet, Inc.

See all devices from Biomet, Inc. →