1525717
by Elekta Inc.
1525717 is a Class II device made by Elekta Inc., cleared under FDA 510(k) K051932 on 12 Aug 2005 and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 05060191071017), active as of 22 Oct 2026. 13 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number05060191071017 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK051932 | Date12 Aug 2005 | StatusActive |
Official records
- AccessGUDID record 05060191071017
- FDA 510(k) database K051932
Across registries
- FDA 510(k) K051932 decided 12 Aug 2005 Active
- FDA GUDID UDI-DI 05060191071017 listed 24 Sep 2016 Listed
Recalls
-
- Terminated Z-2840-2018
There is a potential for an uncontrolled extension of iViewGT / XVI detector arms.
-
- Terminated Z-2721-2016
There have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
-
- Terminated Z-1088-2015
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
-
- Terminated Z-1061-2014
Potential collision risk when using XVI and external beam shaping devices.
-
- Terminated Z-1065-2014
Customers are resetting the default values outside recommended factory settings.
-
- Terminated Z-0931-2013
The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
-
- Terminated Z-2405-2012
There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
-
- Terminated Z-3079-2011
A small number of the Modular Reverse Diodes (D1A), a component in the equipment room of the Elekta Linear Accelerator, have ruptured.
-
- Terminated Z-0305-2012
If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero value.
-
- Terminated Z-1252-2011
Flexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180.
-
- Terminated Z-1723-2011
If another patient is selected on the imaging system (XVI) release 4.5) during the transmission of end of treatment data from Desktop Pro R7.01 to R&V system, the XVI information is prioritized and the end of treatment data is never received by the R&V system.
-
- Terminated Z-0350-2011
It is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode.
-
- Terminated Z-1405-2010
It is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing.
-
- Terminated Z-1411-2010
Firm became aware of a failure in the position detection system which could result in incorrect positioning of the detector in the longitudinal or lateral direction.
Timeline
- Recall initiated FDA recalls Z-0305-2012
- Recall initiated FDA recalls Z-3079-2011
- Recall initiated FDA recalls Z-2405-2012
- Recall initiated FDA recalls Z-0931-2013
- Recall initiated FDA recalls Z-1065-2014
- Recall initiated FDA recalls Z-1061-2014
- Recall initiated FDA recalls Z-1088-2015
- Recall initiated FDA recalls Z-2721-2016
- Listed FDA GUDID 05060191071017
- Recall initiated FDA recalls Z-2840-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Linear accelerator system
- Model / reference
- 1525717
- Primary DI / UDI-DI
- 05060191071017
- 510(k) Number
- FDA product code
- IYE
- GMDN term
- Linear accelerator system
- Device class
- Class II
- Regulation number
- 892.5050
- GUDID distribution status
- In Commercial Distribution