Oec 9800 E/Cv+ Digital Mobile Imaging System

by Ge Oec Medical Systems

510(k) K024012

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

20 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK024012 Date20 Mar 2003ClearedStatusActive

Recalls

20 recalls · FDA

    • Terminated Z-2241-2012 FDA Class II initiated 26 Apr 2010 on shared 510(k) K024012

    GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.

    • Terminated Z-2240-2012 FDA Class II initiated 6 Nov 2009

    GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.

    • Terminated Z-2262-2012 FDA Class II initiated 10 Oct 2008 on shared 510(k) K024012

    GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.

  • 2 recalls, same reason

    • Terminated Z-1885-2008 FDA Unclassified (pre-2012) initiated 12 Aug 2008
    • Terminated Z-1884-2008 FDA Unclassified (pre-2012) initiated 12 Aug 2008

    Beam limitation may be non-compliant on some X-ray units.

    • Terminated Z-1705-2008 FDA Unclassified (pre-2012) initiated 15 Apr 2008

    Use of existing four-pedal footswitch on a different machine may cause various operational errors.

    • Terminated Z-2038-2012 FDA Class I initiated 22 Feb 2008

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is defective and subject to early life failure.

    • Terminated Z-2026-2012 FDA Class II initiated 19 Nov 2007

    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

    • Terminated Z-0990-2007 FDA Unclassified (pre-2012) initiated 6 Jul 2007

    The Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.

    • Terminated Z-0849-2007 FDA Unclassified (pre-2012) initiated 12 Jun 2007

    X-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma.

  • 4 recalls, same reason

    • Terminated Z-0677-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007
    • Terminated Z-0681-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007
    • Terminated Z-0680-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007
    • Terminated Z-0682-2007 FDA Unclassified (pre-2012) initiated 1 Mar 2007

    During routine service, primary and/or secondary collimators may not have been properly reinstalled.

  • 4 recalls, same reason

    • Terminated Z-0556-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007
    • Terminated Z-0558-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007
    • Terminated Z-0559-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007
    • Terminated Z-0557-2007 FDA Unclassified (pre-2012) initiated 9 Feb 2007

    During routine service, a cover may have been installed without the required proper lead shielding.

    • Terminated Z-1302-06 FDA Unclassified (pre-2012) initiated 16 Aug 2006

    X-ray systems could provide output which exceeds the 20 R/minute limit.

    • Terminated Z-0001-2007 FDA Unclassified (pre-2012) initiated 31 May 2006

    Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JAA
Device Class
Class II
Regulation
21 CFR Part 892.1650
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from GE Oec Medical Systems, Inc

See all devices from GE Oec Medical Systems, Inc →