Oec 9800 E/Cv+ Digital Mobile Imaging System
510(k) K024012
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK024012 | Date20 Mar 2003 | StatusActive |
Recalls
-
- Terminated Z-2241-2012
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
-
- Terminated Z-2240-2012
GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.
-
- Terminated Z-2262-2012
GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
-
- Terminated Z-1885-2008
- Terminated Z-1884-2008
Beam limitation may be non-compliant on some X-ray units.
-
- Terminated Z-1705-2008
Use of existing four-pedal footswitch on a different machine may cause various operational errors.
-
- Terminated Z-2038-2012
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is defective and subject to early life failure.
-
- Terminated Z-2026-2012
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.
-
- Terminated Z-0990-2007
The Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.
-
- Terminated Z-0849-2007
X-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma.
-
- Terminated Z-0677-2007
- Terminated Z-0681-2007
- Terminated Z-0680-2007
- Terminated Z-0682-2007
During routine service, primary and/or secondary collimators may not have been properly reinstalled.
-
- Terminated Z-0556-2007
- Terminated Z-0558-2007
- Terminated Z-0559-2007
- Terminated Z-0557-2007
During routine service, a cover may have been installed without the required proper lead shielding.
-
- Terminated Z-1302-06
X-ray systems could provide output which exceeds the 20 R/minute limit.
-
- Terminated Z-0001-2007
Weld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JAA
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent