Skip to content

Optima Coil System OPTI0101HSS10

by Balt USA LLC

Optima Coil System OPTI0101HSS10 is a neurovascular embolization coil (FDA Class II; FDA product code HCG, Device, Neurovascular Embolization) made by Balt USA LLC, cleared under FDA 510(k) K172390 on 18 Feb 2018 and listed in FDA GUDID since 1 Sep 2018 (UDI-DI 00818053022081), active as of 23 Oct 2026. 4 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

510(k) K172390

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00818053022081 UDI-DI Date1 Sep 2018ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK172390 Date18 Feb 2018DecidedStatusActive

Official records

From the FDA record

Device description
The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller.
MRI safety
MR Conditional
Natural rubber latex
No latex statement on the label
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K172390 decided 18 Feb 2018 Active
  • FDA GUDID UDI-DI 00818053022081 listed 1 Sep 2018 Listed

Recalls

4 recalls · 4 ongoing · FDA

    • Ongoing Z-1988-2025 FDA Class II initiated 8 May 2025

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    • Ongoing Z-1041-2024 FDA Class II initiated 3 Jan 2024

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    • Ongoing Z-0103-2023 FDA Class II initiated 7 Sep 2022

    Product pouch label does not match up with carton label

    • Ongoing Z-1680-2022 FDA Class II initiated 26 Jul 2022

    Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.

Timeline

6 dated events

  1. Cleared FDA 510(k) K172390
  2. Listed FDA GUDID 00818053022081
  3. Recall initiated FDA recalls Z-1680-2022
  4. Recall initiated FDA recalls Z-0103-2023
  5. Recall initiated FDA recalls Z-1041-2024
  6. Recall initiated FDA recalls Z-1988-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Optima Coil System
Generic name
Neurovascular embolization coil
Model / Ref / Catalog
OPTI0101HSS10
Primary DI / UDI-DI
00818053022081
510(k) Number
FDA product code
HCG
GMDN term
Neurovascular embolization coil
Device class
Class II
Regulation number
882.5950
GUDID distribution status
In Commercial Distribution
MRI safety
MR Conditional

Similar devices

Other devices filed under FDA product code HCG (Device, Neurovascular Embolization).

Other devices from Balt USA LLC

See all devices from Balt USA LLC →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (9, all answered No)