Skip to content

Optima PET/CT 560

by GE Medical Systems, LLC

510(k) K081496

Optima PET/CT 560 is a medical device (FDA Class II; FDA product code KPS, System, Tomography, Computed, Emission) made by GE Medical Systems, LLC, cleared under FDA 510(k) K081496 on 12 Jun 2008, active as of 22 Oct 2026. 6 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK081496 Date12 Jun 2008ClearedStatusActive

Official records

From the FDA record

Received by the FDA
29 May 2008
FDA decision
12 Jun 2008 (Substantially Equivalent)
FDA review time
14 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K081496 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

7 recalls · 2 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1747-2019 FDA Class II initiated 12 Feb 2019
    • Ongoing Z-1745-2019 FDA Class II initiated 12 Feb 2019

    GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure

    • Terminated Z-2190-2017 FDA Class II initiated 19 May 2017

    It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.

    • Terminated Z-1051-2016 FDA Class II initiated 10 Feb 2016

    GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.

    • Terminated Z-0981-2014 FDA Class II initiated 20 Nov 2013

    GE Healthcare has recently become aware of a potential safety issue due to scatter overcorrection associated with the List Mode Replay feature of your Discovery 600, 610, 710 and Discovery 690.

    • Terminated Z-0024-2014 FDA Class II initiated 26 Mar 2013

    GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID

    • Terminated Z-1188-2013 FDA Class II initiated 25 Jul 2011 on shared 510(k) K081496

    Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

8 dated events

  1. Cleared FDA 510(k) K081496
  2. Recall initiated FDA recalls Z-1188-2013
  3. Recall initiated FDA recalls Z-0024-2014
  4. Recall initiated FDA recalls Z-0981-2014
  5. Recall initiated FDA recalls Z-1051-2016
  6. Recall initiated FDA recalls Z-2190-2017
  7. Recall initiated FDA recalls Z-1747-2019
  8. Recall initiated FDA recalls Z-1745-2019

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
KPS
Device class
Class II
Regulation
21 CFR Part 892.1200
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).

Other devices from GE Medical Systems, LLC