Somatom Emotion 6 CT Systems
by Siemens Medical Systems, Inc.
510(k) K023687
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK023687 | Date22 Nov 2002 | StatusActive |
Recalls
-
- Terminated Z-2472-2018
- Terminated Z-2473-2018
A potential risk of unnecessary radiation exposure due to a software issue
-
- Terminated Z-1300-2016
Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio Pro" and "Cardio Retro".
-
- Terminated Z-1721-2014
The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.
-
- Terminated Z-2452-2008
Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.
-
- Terminated Z-0157-05
Calcium Scoring feature on this devices is not closing properly after patient's exams.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JAK
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent