Uroskop Omnia
by Siemens Medical Solutions USA, Inc.
510(k) K101491
Uroskop Omnia is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K101491 on 30 Jun 2010, active as of 22 Oct 2026. 7 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK101491 | Date30 Jun 2010 | StatusActive |
Official records
- FDA 510(k) database K101491
Recalls
-
- Ongoing Z-1982-2025
- Ongoing Z-1981-2025
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
-
- Terminated Z-0805-2021
After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time.
-
- Terminated Z-0984-2019
The swivel arm holding the monitors may become dislodged from the carrying arm.
-
- Terminated Z-1670-2015
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
-
- Terminated Z-0004-2015
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
-
- Terminated Z-2292-2014
It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. Thi
-
- Terminated Z-0140-2014
When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.
-
- Terminated Z-0374-2014
A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.
-
- Terminated Z-1845-2014
When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.
-
- Terminated Z-0016-2014
Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition. Additionally, it is not possible for the user to stop the radiation manually; ho
-
- Terminated Z-1159-2012
Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.
Timeline
- Recall initiated FDA recalls Z-1845-2014
- Recall initiated FDA recalls Z-0140-2014
- Recall initiated FDA recalls Z-0374-2014
- Recall initiated FDA recalls Z-2292-2014
- Recall initiated FDA recalls Z-0004-2015
- Recall initiated FDA recalls Z-1670-2015
- Recall initiated FDA recalls Z-0984-2019
- Recall initiated FDA recalls Z-0805-2021
- Recall initiated FDA recalls Z-1982-2025
- Recall initiated FDA recalls Z-1981-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAA
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JAA (System, X-Ray, Fluoroscopic, Image-Intensified).
- Custom Pack 08-CC163 by Advance Medical Designs, Inc.
- DR 800 with DSA, DR 800 by Agfa N.V.
- UNBRANDED 5490C by Advance Medical Designs, Inc.
- IMAGE INTENSIFIER TUBE W/HOUSING & by Varian Assoc., Inc.
- MEDRAD FLUOROSCOPIC VIDEO ENHANCEMENT SYSTEM by Medrad, Inc.
- MODEL B-200 C-ARM & MODEL C-300 CARDIAC TABLE by Omega Medical Imaging, LLC
- NEXUS DRF DIGITAL X-RAY IMAGING SYSTEM (WITH PAXSCAN 4343CB) by Varian Medical Systems, X-Ray Products-Infimed
- SIEMEN'S SIREMOBILE II C-ARM by Electromek Diagnostic Systems, Inc.
- Platinum dRF Imaging System by Apelem-Dms Group
- FLUORO AUTOMATIC BRIGHTNESS SYS by Keystone X-Ray, Inc.