Vanguard 360 Instrument 32-360656
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00880304496279 | Date22 Sep 2016 | StatusListed |
Recalls
-
- Terminated Z-0723-2021
Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure
-
- Terminated Z-3249-2018
- Terminated Z-3250-2018
The implant only has one bolt in the sterile package but it should have had two.
-
- Terminated Z-0003-2017
The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be visible, but there are radial cuts into the hex flat. Therefore, when the bolt is inserted into the augment, the screwdriver is unable to catch on the hex geometry, or the screwdriver strips the remaining hex geometry, and the bolt is unable to be fully tightened. The bolt stripping or the bolt not being fully tightened will be noticed by the surgeon intraoperatively when he or she is unable to screw the augment bolts into the augment. This possible event can result in an intraoperative delay resulting in increased time under anesthesia.
-
- Terminated Z-2242-2014
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Vanguard 360 Instrument
- Generic Name
- Knee femur prosthesis trial, reusable
- Model / Ref / Catalog
- 32-360656
- Primary DI / UDI-DI
- 00880304496279
- 510(k) Number
- FDA Product Code
- JWH, MBV
- GMDN Term
- Knee femur prosthesis trial, reusable
- Device Class
- Class II
- Regulation Number
- 888.3560
- GUDID distribution status
- Not in Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
Other devices from Biomet Orthopedics, Inc.
Properties
Other properties (12)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No