Vanguard® 360 Instrument 32-360646
Ref 32-360646
Active
Valid until
Not listed
FDA recalls
None ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number889024BCZ2L 00887868484056 | Date30 Oct 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00887868484056 | Date22 Jun 2022 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Terminated Z-0723-2021
Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure
-
- Terminated Z-3249-2018
- Terminated Z-3250-2018
The implant only has one bolt in the sterile package but it should have had two.
-
- Terminated Z-0003-2017
The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result, the hex geometry can still be visible, but there are radial cuts into the hex flat. Therefore, when the bolt is inserted into the augment, the screwdriver is unable to catch on the hex geometry, or the screwdriver strips the remaining hex geometry, and the bolt is unable to be fully tightened. The bolt stripping or the bolt not being fully tightened will be noticed by the surgeon intraoperatively when he or she is unable to screw the augment bolts into the augment. This possible event can result in an intraoperative delay resulting in increased time under anesthesia.
-
- Terminated Z-2242-2014
The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade Name
- Vanguard® 360 Instrument
- Generic Name
- Knee femur prosthesis trial, reusable
- Device Name
- Vanguard 360 Straight Stem Extension Trials
- Model / Ref / Catalog
- 32-360646
- Primary DI / UDI-DI
- 00887868484056
- Basic UDI-DI
- 889024BCZ2L
- Direct Marketing DI
- 00887868484056
- 510(k) Number
- FDA Product Code
- JWH, MBV
- EMDN Code
- L091199 Orthopaedic prosthetics instruments, reusable - other
- GMDN Term
- Knee femur prosthesis trial, reusable
- Device Class
- Class IIa EUDAMEDClass II GUDID
- Regulation Number
- 888.3560
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class IIa | Class II |
Other devices from Biomet Orthopedics, Inc.
Properties
Other properties (22)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No