Skip to content

Z-800 Infusion System

by Zyno Medical, LLC

510(k) K130690

Z-800 Infusion System is a medical device (FDA Class II; FDA product code FRN, Pump, Infusion) made by Zyno Medical, LLC, cleared under FDA 510(k) K130690 on 12 Jun 2013, active as of 22 Oct 2026. 9 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

9 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK130690 Date12 Jun 2013ClearedStatusActive

Official records

From the FDA record

Received by the FDA
14 Mar 2013
FDA decision
12 Jun 2013 (Substantially Equivalent)
FDA review time
90 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K130690 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

9 recalls · 9 ongoing · FDA

  • 4 recalls, same reason

    • Ongoing Z-1865-2025 FDA Class I initiated 7 May 2025
    • Ongoing Z-1866-2025 FDA Class I initiated 7 May 2025
    • Ongoing Z-1867-2025 FDA Class I initiated 7 May 2025
    • Ongoing Z-1868-2025 FDA Class I initiated 7 May 2025

    Unreleased software versions were installed on distributed devices without verification or validation.

    • Ongoing Z-0506-2025 FDA Class II initiated 18 Oct 2024

    Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

    • Ongoing Z-0523-2025 FDA Class III initiated 30 Sep 2024

    For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.

    • Ongoing Z-0005-2025 FDA Class I initiated 13 Sep 2024

    There is a defect in the air-in-line software algorithm.

    • Ongoing Z-2263-2024 FDA Class II initiated 4 Jun 2024

    Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.

    • Ongoing Z-1183-2024 FDA Class II initiated 20 Dec 2023

    When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death

Timeline

10 dated events

  1. Cleared FDA 510(k) K130690
  2. Recall initiated FDA recalls Z-1183-2024
  3. Recall initiated FDA recalls Z-2263-2024
  4. Recall initiated FDA recalls Z-0005-2025
  5. Recall initiated FDA recalls Z-0523-2025
  6. Recall initiated FDA recalls Z-0506-2025
  7. Recall initiated FDA recalls Z-1865-2025
  8. Recall initiated FDA recalls Z-1866-2025
  9. Recall initiated FDA recalls Z-1867-2025
  10. Recall initiated FDA recalls Z-1868-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
FRN
Device class
Class II
Regulation
21 CFR Part 880.5725
Advisory committee
General Hospital
FDA decision
SESE Substantially Equivalent

Devices cleared under K130690

4 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K130690
UDI-DISize and differences
Z-800
Zyno Medical LLC.
Z-800 / Z-80000814371020006 —On the market
Z-800F
Zyno Medical LLC.
Z-800F / Z-800F00814371020013 —On the market
Z-800W
Zyno Medical LLC.
Z-800W / Z-800W00814371020020 —On the market
Z-800WF
Zyno Medical LLC.
Z-800WF / Z-800WF00814371020037 —On the market

Similar devices

Other devices filed under FDA product code FRN (Pump, Infusion).

Other devices from Zyno Medical LLC.

  • Zyno B2-70071-P

    Class II
  • Zyno A2-80071-PF

    Class II
  • Zyno B2-70070

    Class II
  • Z-800 Open recall

    Class II

See all devices from Zyno Medical LLC. →