Z-800F
Z-800F is a bedside infusion pump (FDA Class II; FDA product code FRN, Pump, Infusion) made by Zyno Medical LLC., cleared under FDA 510(k) K130690 on 12 Jun 2013 and listed in FDA GUDID since 13 Oct 2015 (UDI-DI 00814371020013), active as of 21 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
9 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00814371020013 | Date13 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK130690 | Date12 Jun 2013 | StatusActive |
Official records
- AccessGUDID record 00814371020013
- FDA 510(k) database K130690
From the FDA record
- Device description
- Large volume infusion pump
- MRI safety
- MR Conditional
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K130690 decided 12 Jun 2013 Active
- FDA GUDID UDI-DI 00814371020013 listed 13 Oct 2015 Listed
Recalls
-
- Ongoing Z-1866-2025
- Ongoing Z-1865-2025
- Ongoing Z-1867-2025
- Ongoing Z-1868-2025
Unreleased software versions were installed on distributed devices without verification or validation.
-
- Ongoing Z-0506-2025
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
-
- Ongoing Z-0523-2025
For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
-
- Ongoing Z-0005-2025
There is a defect in the air-in-line software algorithm.
-
- Ongoing Z-2263-2024
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
-
- Ongoing Z-1183-2024
When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death
Timeline
- Listed FDA GUDID 00814371020013
- Recall initiated FDA recalls Z-1183-2024
- Recall initiated FDA recalls Z-2263-2024
- Recall initiated FDA recalls Z-0005-2025
- Recall initiated FDA recalls Z-0523-2025
- Recall initiated FDA recalls Z-0506-2025
- Recall initiated FDA recalls Z-1866-2025
- Recall initiated FDA recalls Z-1865-2025
- Recall initiated FDA recalls Z-1867-2025
- Recall initiated FDA recalls Z-1868-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Generic name
- Bedside infusion pump, single-channel
- Model / Ref / Catalog
- Z-800F
- Primary DI / UDI-DI
- 00814371020013
- 510(k) Number
- FDA product code
- FRN
- GMDN term
- Bedside infusion pump, single-channel
- Device class
- Class II
- Regulation number
- 880.5725
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Similar devices
Other devices filed under FDA product code FRN (Pump, Infusion).
- FLUID MANAGEMENT SYSTEM (MODIFIED) by HAEMONETICS CORPORATION
- MEDTRONIC MICROMED 407C INFUSION PUMP, MODEL MMT-407C by MEDTRONIC MINIMED, INC.
- MEDTRONIC MINIMED MMT-407C AMBULATORY INFUSION PUMP, MODEL MMT-407C by MEDTRONIC MINIMED, INC.
- RATEMINDER V INFUSION PUMP W/SECONDARY CAPABILITY by Critikon Company, LLC
- INFUSION DYNAMICS (ID) POWER INFUSER, MODEL M100B-3A WITH BLOOD CARTRIDGE by Infusion Dynamics, Inc.
- WINDSOR MEDICAL MODEL 50-00002 by Windsor Medical, Inc.
- AVI MICRO 885 DUAL CHANNEL INFUSION PUMP by 3M Avi, Inc.
- RMS Precision Flow Rate TubingTM F0.5 by REPRO MED SYSTEMS, INC.
- IVAC(R) STAR-FLOW(TM) VOLUMETRIC PUMP - MODEL 580 by Y
- PROVIDER ANNE INFUSER by Abbott Laboratories