Balanced Knee System 163-5611
Listed
Valid until
Not applicable
FDA recalls
2 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00822409041329 | Date28 May 2019 | StatusListed |
Recalls
-
- Ongoing Z-1903-2020
- Ongoing Z-1904-2020
The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.
-
- Terminated Z-2255-2014
- Terminated Z-2254-2014
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
-
- Terminated Z-1610-2014
A femoral component containing pegs was found in a box for the pegless version.
-
- Terminated Z-1993-2009
Erroneous lot number on 3 of 28 units of Balanced Knee System. Three units from Lot Number 0055245 were mixed in with the recalled lot (0054891).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Balanced Knee System
- Generic Name
- Tibial insert
- Model / Reference
- 163-5611
- Primary DI / UDI-DI
- 00822409041329
- 510(k) Number
- FDA Product Code
- JWH
- GMDN Term
- Tibial insert
- Device Class
- Class II
- Regulation Number
- 888.3560
- GUDID distribution status
- In Commercial Distribution
- Size
- Height — 11 Millimeter
Other devices from Ortho Development Corp.
Properties
Other properties (8)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No