Cobas 6000 Series System
510(k) K060373
Cobas 6000 Series System is a medical device (FDA Class II; FDA product code CEM, Electrode, Ion Specific, Potassium) made by Roche Diagnostics, cleared under FDA 510(k) K060373 on 13 Mar 2006, active as of 22 Oct 2026.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK060373 | Date13 Mar 2006 | StatusActive |
Official records
- FDA 510(k) database K060373
From the FDA record
- Received by the FDA
- 13 Feb 2006
- FDA decision
- 13 Mar 2006 (Substantially Equivalent)
- FDA review time
- 28 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-2476-2019
Quality issue with high pressure solenoid valves may cause inaccurate results.
-
- Terminated Z-1268-2018
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 sample tubes without tube adapters.
-
- Terminated Z-0909-2018
A warning is being added to the cobas 6000 MODULAR Series System - c501 and the cobas 8000 MODULAR Analyzer Series - c502 Operator Manuals to prevent operator injury during maintenance.
-
- Terminated Z-0378-2018
Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.
-
- Terminated Z-2049-2017
Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.
-
- Terminated Z-0096-2017
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
-
- Terminated Z-1791-2014
- Terminated Z-1792-2014
Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
-
- Terminated Z-0542-2009
Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
-
- Terminated Z-0576-2008
Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.
Timeline
- Recall initiated FDA recalls Z-0576-2008
- Recall initiated FDA recalls Z-0542-2009
- Recall initiated FDA recalls Z-1791-2014
- Recall initiated FDA recalls Z-1792-2014
- Recall initiated FDA recalls Z-0096-2017
- Recall initiated FDA recalls Z-2049-2017
- Recall initiated FDA recalls Z-0378-2018
- Recall initiated FDA recalls Z-0909-2018
- Recall initiated FDA recalls Z-1268-2018
- Recall initiated FDA recalls Z-2476-2019
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CEM
- Device class
- Class II
- Regulation
- 21 CFR Part 862.1600
- Advisory committee
- Clinical Chemistry
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K060373
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Chloride electrode 03246353001 | 03246353001 / 03246353001 | 07613336227787 | On the market | |
| Chloride Electrode 3246353001 | 3246353001 / 3246353001 | 08430215043561 | On the market | |
| cobas 6000 c501 module 7790198001 | 7790198001 / 7790198001 | 04015630946402 | On the market | |
| cobas 6000 Core Unit 07354436001 | 07354436001 / 07354436001 | 04015630938551 | On the market | |
| cobas 6000 core unit 4745868001 | 4745868001 / 4745868001 | 04015630924691 | On the market | |
| cobas c 311 4826876001 | 04826876001 / 4826876001 | 04015630925551 | On the market | |
| cobas c 501 Module 4745914001 | 4745914001 / 4745914001 | 04015630924707 | No longer on the market | |
| cobas e 601 for the cobas 6000 e 601 UL configuration 07354452001 | 07354452001 / 07354452001 | 04015630938568 | On the market | |
| cobas e 601 Immunoassay Analyzer 04745922001 | 04745922001 / 04745922001 | 04015630924714 | No longer on the market | |
| ISE Reference Electrolyte 11360981216 | 11360981216 / 11360981216 | 04015630902682 | On the market | |
| no MS (ISE Reference Electrolyte) 20738085122 | 20738085122 / 20738085122 | 04015630915064 | On the market | |
| OPI Opiates 3002000 20767158122 | 20767158122 / 20767158122 | 04015630996094 | On the market | |
| Opiates II 11986082216 | 11986082216 / 11986082216 | 04015630993284 | No longer on the market | |
| Potassium Electrode 10825441001 | 10825441001 / 10825441001 | 08430215011539 | On the market | |
| Potassium electrode 10825441001 | 10825441001 / 10825441001 | 07613336227770 | On the market | |
| Reference electrode 03149501001 | 03149501001 / 03149501001 | 07613336227794 | On the market | |
| Sodium electrode 10825468001 | 10825468001 / 10825468001 | 08430215011546 | On the market | |
| Sodium electrode 10825468001 | 10825468001 / 10825468001 | 07613336227763 | On the market | |
| Tina-quant Complement C3c Ver.2 08105537190 | 08105537190 / 08105537190 | 07613336133590 | On the market |
Similar devices
Other devices filed under FDA product code CEM (Electrode, Ion Specific, Potassium).
- Dimension® QuikLYTE® Integrated Multisensor Diluent Check 10444872 S640 by Siemens Healthcare Diagnostics Inc.
- SKD ELECTROCHEMICAL SYSTEM (ECS) by Smithkline Diagnostics, Inc.
- Disposable Waste Container D513 REF 905-802 by Radiometer Medical ApS
- ESCHWEILER COMBI-ANALYSER MTA-500 by U.S. Science, Inc.
- ABL90 FLEX SERIES by Radiometer Medical ApS
- POTASSIUM ELECTRODE TIP ASSAMBLY-K- by Hi Chem, Inc.
- ALCYON 300I (WITH ISE MODULE) ANALYZER by Abbott Laboratories
- SENTECH'S ULAB 1000 CLINICAL CHEM. by Sen Tech Medical Corp.
- POSTASSIUM BUFFER SOL. & WASH SOLUTION by Panmed, Inc.
- ILEX PROMPT CLINICAL CHEMISTRY ANALYER by Ilex Corp.