Skip to content

Cobas 6000 Series System

by Roche Diagnostics

510(k) K060373

Cobas 6000 Series System is a medical device (FDA Class II; FDA product code CEM, Electrode, Ion Specific, Potassium) made by Roche Diagnostics, cleared under FDA 510(k) K060373 on 13 Mar 2006, active as of 22 Oct 2026.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

10 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK060373 Date13 Mar 2006ClearedStatusActive

Official records

From the FDA record

Received by the FDA
13 Feb 2006
FDA decision
13 Mar 2006 (Substantially Equivalent)
FDA review time
28 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K060373 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

10 recalls · FDA

    • Terminated Z-2476-2019 FDA Class II initiated 18 Jul 2019 on shared 510(k) K060373

    Quality issue with high pressure solenoid valves may cause inaccurate results.

    • Terminated Z-1268-2018 FDA Class II initiated 18 Dec 2017 on shared 510(k) K060373

    Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 sample tubes without tube adapters.

    • Terminated Z-0909-2018 FDA Class II initiated 18 Jul 2017 on shared 510(k) K060373

    A warning is being added to the cobas 6000 MODULAR Series System - c501 and the cobas 8000 MODULAR Analyzer Series - c502 Operator Manuals to prevent operator injury during maintenance.

    • Terminated Z-0378-2018 FDA Class II initiated 29 Jun 2017 on shared 510(k) K060373

    Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.

    • Terminated Z-2049-2017 FDA Class II initiated 21 Mar 2017 on shared 510(k) K060373

    Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.

    • Terminated Z-0096-2017 FDA Class II initiated 15 Jul 2016 on shared 510(k) K060373

    Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

  • 2 recalls, same reason

    • Terminated Z-1791-2014 FDA Class II initiated 21 May 2014 on shared 510(k) K060373
    • Terminated Z-1792-2014 FDA Class II initiated 21 May 2014 on shared 510(k) K060373

    Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.

    • Terminated Z-0542-2009 FDA Unclassified (pre-2012) initiated 15 Sep 2008 on shared 510(k) K060373

    Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.

    • Terminated Z-0576-2008 FDA Unclassified (pre-2012) initiated 4 Dec 2007 on shared 510(k) K060373

    Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

11 dated events

  1. Recall initiated FDA recalls Z-0576-2008
  2. Recall initiated FDA recalls Z-0542-2009
  3. Recall initiated FDA recalls Z-1791-2014
  4. Recall initiated FDA recalls Z-1792-2014
  5. Recall initiated FDA recalls Z-0096-2017
  6. Recall initiated FDA recalls Z-2049-2017
  7. Recall initiated FDA recalls Z-0378-2018
  8. Recall initiated FDA recalls Z-0909-2018
  9. Recall initiated FDA recalls Z-1268-2018
  10. Recall initiated FDA recalls Z-2476-2019

1 earlier event are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
CEM
Device class
Class II
Regulation
21 CFR Part 862.1600
Advisory committee
Clinical Chemistry
FDA decision
SESE Substantially Equivalent

Devices cleared under K060373

19 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K060373
UDI-DISize and differences
Chloride electrode 03246353001
Roche Diagnostics GmbH
03246353001 / 0324635300107613336227787
Prescription: yes; Single use: no
On the market
Chloride Electrode 3246353001
Roche Diagnostics GmbH
3246353001 / 324635300108430215043561
Prescription: yes; Single use: yes
On the market
cobas 6000 c501 module 7790198001
Roche Diagnostics GmbH
7790198001 / 779019800104015630946402
Prescription: yes; Single use: no
On the market
cobas 6000 Core Unit 07354436001
Roche Diagnostics GmbH
07354436001 / 0735443600104015630938551
Prescription: no; Single use: no
On the market
cobas 6000 core unit 4745868001
Roche Diagnostics GmbH
4745868001 / 474586800104015630924691
Prescription: yes; Single use: no
On the market
cobas c 311 4826876001
Roche Diagnostics GmbH
04826876001 / 482687600104015630925551
Prescription: yes; Single use: no
On the market
cobas c 501 Module 4745914001
Roche Diagnostics GmbH
4745914001 / 474591400104015630924707
Prescription: yes; Single use: no
No longer on the market
cobas e 601 for the cobas 6000 e 601 UL configuration 07354452001
Roche Diagnostics GmbH
07354452001 / 0735445200104015630938568
Prescription: no; Single use: no
On the market
cobas e 601 Immunoassay Analyzer 04745922001
Roche Diagnostics GmbH
04745922001 / 0474592200104015630924714
Prescription: no; Single use: no
No longer on the market
ISE Reference Electrolyte 11360981216
Roche Diagnostics GmbH
11360981216 / 1136098121604015630902682
Prescription: yes; Single use: yes
On the market
no MS (ISE Reference Electrolyte) 20738085122
Roche Diagnostics GmbH
20738085122 / 2073808512204015630915064
Prescription: yes; Single use: yes
On the market
OPI Opiates 3002000 20767158122
Roche Diagnostics GmbH
20767158122 / 2076715812204015630996094
Prescription: yes; Single use: yes
On the market
Opiates II 11986082216
Roche Diagnostics GmbH
11986082216 / 1198608221604015630993284
Prescription: yes; Single use: yes
No longer on the market
Potassium Electrode 10825441001
Roche Diagnostics GmbH
10825441001 / 1082544100108430215011539
Prescription: yes; Single use: yes
On the market
Potassium electrode 10825441001
Roche Diagnostics GmbH
10825441001 / 1082544100107613336227770
Prescription: yes; Single use: no
On the market
Reference electrode 03149501001
Roche Diagnostics GmbH
03149501001 / 0314950100107613336227794
Prescription: yes; Single use: no
On the market
Sodium electrode 10825468001
Roche Diagnostics GmbH
10825468001 / 1082546800108430215011546
Prescription: yes; Single use: yes
On the market
Sodium electrode 10825468001
Roche Diagnostics GmbH
10825468001 / 1082546800107613336227763
Prescription: yes; Single use: no
On the market
Tina-quant Complement C3c Ver.2 08105537190
Roche Diagnostics GmbH
08105537190 / 0810553719007613336133590
Prescription: yes; Single use: yes
On the market

Similar devices

Other devices filed under FDA product code CEM (Electrode, Ion Specific, Potassium).

Other devices from Roche Diagnostics Corp.

See all devices from Roche Diagnostics Corp. →