Homechoice Apd Set 5C4469C
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number07332414208479 | Date16 Oct 2025 | StatusListed |
Recalls
-
- Terminated Z-2132-2012
Baxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.
-
- Terminated Z-0946-2008
Unrecorded Infusion; If the system is powered down or a power failure occurs during a fill cycle, the HomeChoice system may not record the last 1-4 pump strokes infused into the patient when the power is restored. This situation may create a potential unrecorded delivery of approximately 15-60 mL of fluid being infused into the patient.
-
- Terminated Z-1018-05
The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an electric shock hazard.
-
- Terminated Z-1012-04
A software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the volume is made at the Verify I-DRAIN:XXXXML prompt.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Homechoice Apd Set
- Generic Name
- Peritoneal dialysis system tubing set
- Model / Ref / Catalog
- 5C4469C
- Primary DI / UDI-DI
- 07332414208479
- 510(k) Number
- FDA Product Code
- FKX
- GMDN Term
- Peritoneal dialysis system tubing set
- Device Class
- Class II
- Regulation Number
- 876.5630
- GUDID distribution status
- In Commercial Distribution
Other devices from Vantive US Healthcare, LLC
Properties
Other properties (8)
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No