microTargeting™ Insertion Tube 66-PT(IT2)-00779
by Fhc, Inc. GUDID
Listed Open recall
Regulatory status
Recall ongoing. 1 ongoing FDA recall names this device. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00873263007792 | Date8 May 2026 | StatusListed |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-0312-2025
A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
-
- Terminated Z-2346-2018
- Terminated Z-2355-2018
- Terminated Z-2348-2018
- Terminated Z-2356-2018
- Terminated Z-2347-2018
- Terminated Z-2345-2018
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
-
- Terminated Z-0895-05
Potential for non-sterile pin to contaminate sterile field
Specification
- Trade Name
- microTargeting™ Insertion Tube
- Generic Name
- Intracerebral cannula, nonimplantable, reusable
- Model / Ref / Catalog
- 66-PT(IT2)-00779, 66-PT(IT2)
- Primary DI / UDI-DI
- 00873263007792
- 510(k) Number
- FDA Product Code
- HAW
- GMDN Term
- Intracerebral cannula, nonimplantable, reusable
- Device Class
- Class II
- Regulation Number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
- Sterilization Method
- Moist Heat or Steam Sterilization
Other devices from Fhc, Inc.
Properties
Other properties (11)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No