Skip to content

Monaco

by Impac Medical Systems, Inc. 2 variants

510(k) K132971

Monaco is a medical device (FDA Class II; FDA product code MUJ, System, Planning, Radiation Therapy Treatment) made by Impac Medical Systems, Inc., cleared under FDA 510(k) K132971 on 20 Nov 2013, active as of 22 Oct 2026. 4 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK132971 Date20 Nov 2013ClearedStatusActive

Official records

From the FDA record

Received by the FDA
23 Sep 2013
FDA decision
20 Nov 2013 (Substantially Equivalent)
FDA review time
58 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K132971 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

4 recalls · FDA

    • Terminated Z-1009-2017 FDA Class II initiated 21 Dec 2016

    Incorrect Enhanced Dynamic Wedge (EDW) or Virtual Wedge (VW) Calculations.

    • Terminated Z-0668-2017 FDA Class II initiated 17 Nov 2016

    Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.

    • Terminated Z-1237-2015 FDA Class II initiated 19 Dec 2014

    Incorrect CT to ED when using the Monaco Image Statistics Tool.

    • Terminated Z-2581-2014 FDA Class II initiated 18 Aug 2014

    Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.

Timeline

5 dated events

  1. Cleared FDA 510(k) K132971
  2. Recall initiated FDA recalls Z-2581-2014
  3. Recall initiated FDA recalls Z-1237-2015
  4. Recall initiated FDA recalls Z-0668-2017
  5. Recall initiated FDA recalls Z-1009-2017

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
MUJ
Device class
Class II
Regulation
21 CFR Part 892.5050
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).

Other devices from Impac Medical Systems, Inc.

See all devices from Impac Medical Systems, Inc. →