Na 1525340

by Elekta Inc.

Listed FDA GUDID listing, no authorization on record

Valid until

Not applicable

Database listing, no authorization

FDA recalls

None ongoing

14 closed

Documents

None

No authorization documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source GUDID, synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number05060191070911 UDI-DI Date24 Sep 2016ListedStatusListed GUDID: in commercial distribution

Recalls

14 recalls · FDA

    • Terminated Z-2840-2018 FDA Class II initiated 1 Dec 2017

    There is a potential for an uncontrolled extension of iViewGT / XVI detector arms.

    • Terminated Z-2721-2016 FDA Class II initiated 8 Jul 2016

    There have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.

    • Terminated Z-1088-2015 FDA Class II initiated 7 Jan 2015

    Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.

    • Terminated Z-1061-2014 FDA Class II initiated 25 Nov 2013

    Potential collision risk when using XVI and external beam shaping devices.

    • Terminated Z-1065-2014 FDA Class II initiated 28 May 2013

    Customers are resetting the default values outside recommended factory settings.

    • Terminated Z-0931-2013 FDA Class II initiated 19 Dec 2012

    The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.

    • Terminated Z-2405-2012 FDA Class II initiated 6 Aug 2012

    There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.

    • Terminated Z-3079-2011 FDA Unclassified (pre-2012) initiated 6 May 2011 on shared 510(k) K051932

    A small number of the Modular Reverse Diodes (D1A), a component in the equipment room of the Elekta Linear Accelerator, have ruptured.

    • Terminated Z-0305-2012 FDA Unclassified (pre-2012) initiated 9 Mar 2011

    If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero value.

    • Terminated Z-1252-2011 FDA Unclassified (pre-2012) initiated 5 Jan 2011

    Flexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180.

    • Terminated Z-1723-2011 FDA Unclassified (pre-2012) initiated 25 Oct 2010

    If another patient is selected on the imaging system (XVI) release 4.5) during the transmission of end of treatment data from Desktop Pro R7.01 to R&V system, the XVI information is prioritized and the end of treatment data is never received by the R&V system.

    • Terminated Z-0350-2011 FDA Unclassified (pre-2012) initiated 1 Sep 2010

    It is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode.

    • Terminated Z-1405-2010 FDA Unclassified (pre-2012) initiated 12 Dec 2008

    It is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing.

    • Terminated Z-1411-2010 FDA Unclassified (pre-2012) initiated 19 Sep 2007

    Firm became aware of a failure in the position detection system which could result in incorrect positioning of the detector in the longitudinal or lateral direction.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — GUDID listing Listed

Specification

GUDID

Generic Name
Linear accelerator system
Model / Reference
1525340
Primary DI / UDI-DI
05060191070911
510(k) Number
FDA Product Code
IYE
GMDN Term
Linear accelerator system
Device Class
Class II
Regulation Number
892.5050
GUDID distribution status
In Commercial Distribution

Other devices from Elekta Inc.

See all devices from Elekta Inc. →

Properties

  • Prescription Use
  • Serial Number
  • Manufacturing Date

Other properties (11)

  • Sterile: No
  • Single Use: No
  • Implantable: No
  • Contains Latex: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Lot / Batch Number: No
  • Expiration Date: No