Skip to content

S31KP

by Fujifilm Corporation

Ref 04573596212301Model S31KP

S31KP is a device classified as Class IIa in the EU and Class II in the US made by Fujifilm Corporation, registered in EUDAMED on 30 Oct 2026 (Basic UDI-DI B-04573596212301), cleared under FDA 510(k) K220295 on 29 Apr 2022 (6 earlier submissions) and listed in FDA GUDID since 18 Jul 2021 (UDI-DI 04573596212301), active as of 7 Nov 2026. 4 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDD 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Source EUDAMED, GUDID, synced 7 Nov 2026.

Regulatory status

as of registry sync 7 Nov 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDD EUDAMED NumberB-04573596212301 Basic UDI-DI 04573596212301UDI-DIDate30 Oct 2026RegisteredStatusActive
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Abbreviated NumberK220295 Date29 Apr 2022DecidedStatusActive
United StatesGUDID listing GUDID Number04573596212301 UDI-DI Date18 Jul 2021ListedStatusListed GUDID: in commercial distribution

Official records

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI B-04573596212301 registered 30 Oct 2026 Active
  • FDA 510(k) K220295 decided 29 Apr 2022 Active· +6 more in the status table
  • FDA GUDID UDI-DI 04573596212301 listed 18 Jul 2021 Listed

Recalls

4 recalls · 2 ongoing · FDA

    • Ongoing Z-1884-2024 FDA Class II initiated 11 Apr 2024

    AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

    • Ongoing Z-0189-2022 FDA Class II initiated 14 Oct 2021

    Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

    • Terminated Z-1341-2020 FDA Class II initiated 31 Jan 2020

    These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.

    • Terminated Z-2090-2017 FDA Class II initiated 2 May 2017

    The strength of the arm mounting mechanism was insufficient and had to be redesigned.

Timeline

13 dated events

  1. Recall initiated FDA recalls Z-2090-2017
  2. Cleared FDA 510(k) K163505
  3. Cleared FDA 510(k) K191233
  4. Recall initiated FDA recalls Z-1341-2020
  5. Cleared FDA 510(k) K202422
  6. Listed FDA GUDID 04573596212301
  7. Recall initiated FDA recalls Z-0189-2022
  8. Cleared FDA 510(k) K220295
  9. Recall initiated FDA recalls Z-1884-2024
  10. Registered EUDAMED B-04573596212301

3 earlier events are in the sections above.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDD Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
S31KP Probe EUDAMED
S31KP GUDID
Generic name
Surgical ultrasound imaging transducer
Model / reference
S31KP, 04573596212301
Primary DI / UDI-DI
04573596212301
Basic UDI-DI
B-04573596212301
FDA product code
IYO, IYN, ITX
EMDN code
Z11040299 Ultrasound probes – other
GMDN term
Surgical ultrasound imaging transducer
Device class
Class IIa EUDAMED
Class II GUDID
Regulation number
892.1560, 892.1550, 892.1570
EU registration type
Legacy device (under a directive)
EU market of registration
Netherlands
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class IIa
GUDID
Class II

Device class is stated on each registry's own scale: Class IIa in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under ULTRASOUND PROBES – OTHER.

Other devices from Fujifilm Corporation

See all devices from Fujifilm Corporation →

Properties

  • Sterilise before use
  • Prescription use
  • Serial number
  • Manufacturing date
Other properties (19, all answered No)