Trilogy Evo Universal
510(k) K181170
Trilogy Evo Universal is a medical device (FDA Class II; FDA product code CBK, Ventilator, Continuous, Facility Use) made by Respironics, Inc., cleared under FDA 510(k) K181170 on 18 Jul 2019, active as of 22 Oct 2026. 11 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
13 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181170 | Date18 Jul 2019 | StatusActive |
Official records
- FDA 510(k) database K181170
From the FDA record
- Received by the FDA
- 2 May 2018
- FDA decision
- 18 Jul 2019 (Substantially Equivalent)
- FDA review time
- 442 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1617-2026
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
-
- Ongoing Z-1649-2026
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
-
- Ongoing Z-0292-2025
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
-
- Ongoing Z-2943-2024
- Ongoing Z-2944-2024
- Ongoing Z-2942-2024
Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.
-
- Ongoing Z-1504-2024
- Ongoing Z-1505-2024
Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
-
- Ongoing Z-1652-2023
The Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and pulse rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional and hospital settings and non-emergency transport settings; for example, wheelchair. It may be used for both invasive and non-invasive ventilation.
-
- Ongoing Z-1289-2023
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.
-
- Ongoing Z-0493-2022
- Ongoing Z-0494-2022
Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.
-
- Ongoing Z-2290-2021
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Timeline
- Recall initiated FDA recalls Z-1289-2023
- Recall initiated FDA recalls Z-1652-2023
- Recall initiated FDA recalls Z-1504-2024
- Recall initiated FDA recalls Z-1505-2024
- Recall initiated FDA recalls Z-2943-2024
- Recall initiated FDA recalls Z-2944-2024
- Recall initiated FDA recalls Z-2942-2024
- Recall initiated FDA recalls Z-0292-2025
- Recall initiated FDA recalls Z-1617-2026
- Recall initiated FDA recalls Z-1649-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CBK
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory committee
- Anesthesiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K181170
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Trilogy Evo Universal DS2000X11B | DS2000X11B / DS2000X11B | 00606959052000 | — | On the market |
Similar devices
Other devices filed under FDA product code CBK (Ventilator, Continuous, Facility Use).
- AVEA VENTILATOR by CAREFUSION
- AMARA GEL 1090436 by Respironics Inc.
- IVENT 201 by Versamed Medical Systems, Ltd.
- INSPIRAID VA-1 VENTILATOR by Inspir Aid
- OXYLOG 2000 by Drager, Inc.
- MODIFICATION TO I VENT 201 by Versamed Medical Systems, Ltd.
- ENGSTROM CARESTATION, ENGSTROM PRO by Datex-Ohmeda Inc., A Division of Instrumentarium C
- HAMILTON-G5 by Hamilton Medical AG
- Ventoux VC2 VX-08-1-MN1 by FLIGHT MEDICAL INNOVATIONS LTD
- ENGSTROM ERICA II RESPIRATOR by Gambro, Inc.